Vaisala Air Quality Transmitter AQT410

Overview

Vaisala Air Quality Transmitter AQT410 measures the most common gaseous pollutants nitrogen dioxide (NO2), sulphur dioxide (SO2), carbon monoxide (CO) and ozone (O3). The AQT410 measurement performance is based on proprietary advanced algorithms that enable ppb measurements at an affordable price using electrochemical sensors.

AQT410 has been specifically designed for air quality monitoring networks in urban areas, road networks or around industrial sites and transportation hubs. Thanks to its small weight and compact size it is ideally suited for deployment even in large air quality networks.

Applications

  • Urban air quality networks
  • Industrial emission monitoring
  • Safety monitoring
  • Roadside and tunnel monitoring
  • Mobile measurement
  • Building automation
  • Air quality research

Features

  • Measures up to four most common air pollutants NO2, SO2, CO and O3.
  • Intelligent algorithms that compensate for aging and environmental conditions
  • Compact design, easy to deploy in the field
  • Low power consumption (typically 0.5W)
  • Wireless Internet connection with Multi-Observation Gateway
  • RS232 and RS485 interfaces for local connectivity (eg. Modbus support)
  • Easy integration and open API
  • Technical Specifications
    General
    Data protocolsModbus, ASCII
    Serial data interfaceRS-485
    Console interfaceRS-232
    Power and data connectorStandard 8-pin M12 male
    Operating voltage8 — 30 VDC
    Power consumptionTyp. 0.5 W, max. 2 W
    Protection classIP65
    Enclosure materialsAnodized aluminum, stainless steel
    Dimensions125(w) x 125(h) x 128(d) mm
    Weight690 g
    Measurement Specifications
    Temperature range-40 — 85 °C
    Temperature resolution0.1 °C
    Temperature accuracy±0.3 °C, repeatability ±0.1 °C
    Humidity range0 — 100 %RH (non-condensing)
    Humidity resolution0.1 %RH
    Humidity accuracy±2 %RH, repeatability ±0.2 %RH
    Pressure range800 — 1100 mbar
    Pressure resolution1 mbar
    Pressure accuracy<±1 % FS
    Air Quality Measurement Specifications
    Temperature range-30 — 50 °C
    Humidity range15 — 95 %RH (non-condensing)
    Gas Measurement Specifications
    GASRANGEMIN. DETECTIONRESOLUTIONPRECISIONLINEARITYUNIT
    SO20 — 20.005±0.001<±1 % FS<±1 % FSppm
    NO20 — 20.005±0.001<±1 % FS<±1 % FSppm
    CO0 — 100.01±0.01<±2 % FS<±2 % FSppm
    O30 — 20.005±0.01<±3 % FS<±2 % FSppm
    Particle Measurement Specifications
    Particle counter channelsPM2.5 and PM10
    Particle range0.3 — 20 µm (spherical equivalent)
    Response time<60 s
    Sampling interval1 — 1440 minutes
    Sample flow rate0.5 SLM (integrated vacuum pump)
    Unitsµg/m3
    Measurement range for PM2.50 — 2000 µg/m3
    Measurement range for PM100 —5000 µg/m3
    Measurement resolution0.1 µg/m3
    Conformity
    EMCIEC/EN 61326-1, IEC/EN61000-4-2/3/4/5/6, CISPR 22I
    Talent is God given. Be humble. Fame is man-given. Be grateful. Conceit is self-given. Be careful.